Figure 2 - example: disassociation between ECG and arterial pulse.
Welcome to the 14th edition of our biweekly newsletter! In this issue, we’re diving into a fascinating case study that highlights a common discrepancy and provides valuable insights for your daily practice.
Case Study: The Curious Case of the Mismatched Heart Rates
We recently received an intriguing case from a veterinarian. The patient was a 10-year-old, 19.3 kg mixed-breed dog presenting with clinical symptoms of syncope, increased respiratory rate, and ascites. A grade 3 heart murmur was noted. The veterinarian submitted an ECG to our platform, which generated a report with some surprising findings.
The CardioBird ECG report indicated a heart rate of 250 bpm, along with the presence of atrial fibrillation (AFib) with a possible right bundle branch block and a deep S wave (Fig. 1). However, the veterinarian noted a significant difference between this heart rate and the one they had measured using auscultation and a blood pressure monitor. This discrepancy sparked a great conversation with our cardiology team about the nuances of heart rate measurement.
Figure 1 – Abnormal ECG of the patient
Why the Discrepancy? Understanding the Science
Our cardiologists pointed out that auscultation, while a fundamental tool, may not always be a reliable method for measuring heart rate, especially in cases with high heart rates or complex arrhythmias like AFib or ventricular tachycardia (VT). In such conditions, some heartbeats may be too weak to generate a palpable pulse or a strong enough sound to be detected by the human ear. This phenomenon is known as a “pulse deficit.”
Figure 2 – example: disassociation between ECG and arterial pulse.
The same principle applies to blood pressure monitors, which rely on the arterial pulse to determine heart rate. Our cardiologists shared a compelling example image from a monitoring session (fig. 2), showing that the arterial waveform (measured by the blood pressure monitor) was not consistently associated with every electrical heartbeat shown on the ECG. This visual evidence clearly demonstrated that not every electrical impulse from the heart results in a strong enough contraction to create a detectable pulse wave in the arteries. This is a classic sign of a pulse deficit, which is common in conditions like severe AFib.
In this specific case, the dog’s high heart rate and AFib likely led to an incomplete ventricular filling, resulting in some beats being too weak to produce a peripheral pulse. The CardioBird ECG, which measures the heart’s electrical activity directly, captured every single electrical depolarization, thus providing a more accurate and higher heart rate measurement.
Best Practices for Your Clinic
So, what can we learn from this case?
This case serves as a powerful reminder that while technology can sometimes present seemingly contradictory information, understanding the underlying physiological principles allows us to use these tools to their full potential for our patients’ well-being.
Estimated reading time: 4 minutes Welcome to the first installment of our new series,…
Welcome to our new series, "The Turning Point," where we spotlight real-world cases from…
Veterinarian Spotlight #17: Dr. Gayeon An Bio: DVM/ / Jeju National University College…
A common question we receive is why the respiratory rate derived from an ECG…
Estimated reading time: 3.48 minutes A CardioBird Whitepaper for Veterinarians For decades, the electrocardiogram…
Veterinarian Spotlight #16: Dr. Wei-Chih Hsu We are delighted to introduce Dr. Hsu as…